When you receive your Leckey Prone/Supine Stander it will be boxed on a pallet. You should first cut the straps and remove the outer lid and cardboard sleeve. The tray will be packaged separately and should be removed from the box before trying to move the stander. You can then remove the remaining packaging, unlock the castors and roll the stander off the pallet. Please check that you have all the parts you have ordered. The castors should then be locked into position, facing out from the frame at 45 degrees. Some of the components will be wrapped in polythene packaging.
Safety First.
Keep polythene bags away from children.


Multi-tool

A number of adjustments will require the use of the multi-tool, which is supplied with each product.                                                                                                                                          

Charging the battery

Before the product is used for the first time it is recommended to charge the battery fully for at least 12 hours. To charge the battery, plug the adaptor into the mains socket, attach the lead (a) to the handset and switch on the mains power. While charging the charger LED will flash orange and will turn to green when fully charged. When charged switch off mains power, remove the adaptor plug when and disconnect the lead from the battery. The battery charge should be topped up each day for approximately one hour. When charged switch off the mains power, remove the adaptor plug and disconnect the lead from the handset. However if the battery is low in power an audible bleep will sound indicating that it needs to be charged.                                                                               Setting the Angle
Whether used in supine or prone the frame angle can be easily adjusted with the push button control handset. Ensure that the product and user are away from surrounding furnishings to prevent possible collisions when adjusting the angle.
Safety first
Keep polythene bags away from children.

How to Attach the tray in prone standing
The tray is attached to and removed from the frame by sliding the square metal tube into the top of the main spine. Before you insert the tube unlock the locking lever (a), insert the tube and relock the lever. To lock the tray and set its height push the over centre lever (a) down until it touches the metal spine. After use the locking lever need to be fine tuned to ensure secure locking. To do this lift up the lever and turn the nut (b) while holding the lever in a fixed position. Push the lever back down to check if it locks the tube securely. Repeat the process as necessary.
Always use caution when making this adjustment, as fingers could become trapped in the receiving tube while positioning.


How to Attach the tray in supine standing

The Tray is attached to and removed from the frame by the metal brackets located below the chest pad. To attach the tray slide the metal bars into the receiving tubes to the required depth and secure by tightening the knobs (a) on either side.


How to attach the sandals

Each sandal is attached to the hinged footplate individually with a single fixing bolt. Remove the sandals from their packaging and unscrew the plastic knob and washers. Place the sandal on the footplate locating the fixing bolt through the slot, place the rubber washer on first and then the
metal washer. Secure in place with the plastic knob You can flip up the footplate for easier access.
Always use caution when making this adjustment, as fingers could become trapped in the slot while positioning.



How to attach the headsupport in Supine

To attach and height adjust, slide the square tubing into the top of the central spine on the backrest. To lock the head support push the over centre lever (a) down until it touches the metal spine. To remove simply reverse this process. The Whitmyer Pro-Series range of headsupports can also be attached to the frame as detailed above. After use locking lever may need to be fine tuned to ensure secure locking. To do this lift up the lever and turn the nut (b) while holding the lever in a fixed position. Push the lever back down to check if it locks the tube securely. Repeat the process as necessary.
Always use caution when making this adjustment, as fingers could become trapped in the receiving tube while positioning.


How to attach the headsupport in prone

If headsupport is need in prone standing the Leckey Interface Frame (LIF) can be used with Horizon Stander. The LIF is attached to and removed from the frame by the metal brackets located below the chest pad. To attach the slide the metal bars into the receiving tubes to the required depth
and secure by tightening the knobs (a) on either side. Please refer to the specific user manual on the Leckey Interface Frame to safely position
the frame and its accessories. This frame will also accommodate the Whitmyer pro-series range of headsupports.

How to attach Flexible Hip Pads
To attach the flexible guides, loosen the plastic knob at the back of the pad and slide the metal bracket into the slot. Then when set to the desired with retighten the plastic knob. Repeat on the other side.