Daily and Annual Product Inspections

Daily Product Inspection (Therapists, parents & carers)
We recommend that daily visual checks of the equipment are carried out by therapists, carers or parents to ensure the product is safe for use. The recommended daily checks are detailed below.
1. Check that the product is clean and there are no signs of soiling.
2. Check that all fabrics, webbing, Velcro and plastic buckles are in good condition and do not show signs of wear and tear or fraying.
3. Check the sleepform mattress is air tight; by extracting the air and does not loose its shape after the valve has been closed.
4. Check the Sleepform mattress for scratches,scores or punctures. Seal any holes with the repair patches supplied with the product. Holes can be found by visual inspection or by covering the surface of the mattress with soapy water and watching for air bubbles. If a hole has been found clean the area and let it dry. Select a patch and put glue on it and on the area around the hole. Wait for 2 minutes and firmly press the patch over the hole ensuring that no air remains under the patch. The bond will be complete after 24 hours.
DO NOT FULLY IMMERSE THE SLEEPFORM MATTRESS.
 
Annual Product Inspection(Therapist, Technician, Leckey Product Advisor, Dealer)
Leckey recommend that each product should be subject to a detail inspection at least one a year and every time the product is reissued for use. This inspection should be carried out by a technically competent person who has been trained in the use of the product and should include the following checks as a minimum requirement.
1. Check that the product is clean and there are no signs of soiling.
2. Check that all fabrics, webbing, Velcro and plastic buckles are in good condition and do not show signs of wear and tear or fraying.
3. Check that all buckles lock securely when clipped together.
4. Check the sleepform mattress is air tight; by extracting the air and does not loose its shape after the valve has been closed.
5. Check the valve on the Sleepform mattress is working properly.
6. Check the Sleepform mattress for scratches,scores or punctures. Seal any holes with the repair patches supplied with the product. Holes can be found by visual inspection or by covering the surface of the mattress with soapy water and watching for air bubbles. If a hole has been found clean the area and let it dry. Select a patch and put glue on it and on the area around the hole. Wait for 2 minutes and firmly press the patch over the hole ensuring that no air remains under the patch. The bond will be complete after 24 hours.
DO NOT FULLY IMMERSE THE SLEEPFORM MATTRESS.
7. Check the airflow mattress is intact and that the cushion is even and flat and that it has not gathered up within the cover.